Improving the accuracy of gastrointestinal neuroendocrine tumor grading with deep learning
The Ki-67 index is an established prognostic factor in gastrointestinal neuroendocrine tumors (GI-NETs) and defines tumor grade. It is currently estimated by microscopically examining tumor tissue sin
Summary
The Ki-67 index sets the grade of a gastrointestinal neuroendocrine tumour (GI-NET), but the manual method is unreliable. A pathologist picks a hot-spot by eye and counts 500–2,000 cells, or estimates by eye. On a Ki-67-only stain, proliferating non-tumour cells (endothelium, crypts, lymphocytes) are easily counted by mistake.
This UC Davis study uses a synaptophysin/Ki-67 double stain so the software can count only inside synaptophysin-positive tumour, and builds two tools on it:
SKIE finds hot-spots automatically on the whole-slide image and computes the Ki-67 index there.
deep-SKIE is a deep learner that paints a tile-by-tile grade heatmap across the whole tumour.
Because double staining is not routine, a cycle GAN converts ordinary Ki-67-only slides into virtual double stains.
The abstract in the
description:frontmatter is the paper's own, verbatim. The sections below are my own-words digest of the abstract and the open-access full text (CC BY 4.0) in PubMed Central.
Citation
Govind D, Jen KY, Matsukuma K, Gao G, Olson KA, Gui D, et al. Improving the accuracy of gastrointestinal neuroendocrine tumor grading with deep learning. Sci Rep. 2020;10(1):11064. Published 2020 Jul 6. doi: 10.1038/s41598-020-67880-z. PMID: 32632119. PMCID: PMC7338406.
Models: SKIE (Synaptophysin-Ki-67 Index Estimator) and deep-SKIE
Access: open access, CC BY 4.0
Study at a glance
Cases
50 GI-NETs, grade 1 and 2 only (Ki-67 ≥ 20% excluded)
Stains
Synaptophysin (Permanent Red) / Ki-67 (DAB) double stain, with adjacent Ki-67-only and H&E sections; Aperio AT2 at 20×
Gold standard
A fourth pathologist exhaustively ticking off at least 500 tumour cells in one captured hot-spot on the double-stained WSI
Comparators
Three GI pathologists reading the double-stained glass slides at the microscope; ImmunoRatio
SKIE
Registers H&E to the double stain; takes the tumour mask from the red channel; picks five candidate 500 × 500 µm hot-spots; counts Ki-67-negative nuclei on the H&E
deep-SKIE
Inception V3 trained and validated on 42 cases (15,232 tiles), tested on 6 cases (9,436 tiles); four classes: background, non-tumour, G1, G2
Cycle GAN
Trained on about 47,000 unpaired tiles from 42 WSIs; tested on 6 hold-out WSIs
Key findings
Grade agreement. SKIE matched the gold-standard grade in 45 of 50 cases (90%; linear-weighted κ = 0.62), with a Ki-67 index error of 0.84 ± 1.02%. With two technically flawed cases excluded, it matched 45 of 48 (93.8%).
The discordant cases were the algorithm's wins. In the three remaining disagreements, SKIE had chosen a hotter hot-spot. When the gold-standard pathologist counted SKIE's hot-spots, all three were regraded from G1 to G2. Overall, SKIE chose a higher-index hot-spot than the gold-standard pathologist in 36 of 50 cases.
Same field, same answer. On the gold-standard pathologist's own hot-spots, SKIE matched the grade in 48 of 50 cases (96%).
Pathologists vs the gold standard. κ was 0.32, 0.67 and 0.78 for the three microscope readers.
Against ImmunoRatio. SKIE had about 2.5-fold lower error, with the biggest gains where Ki-67-positive non-tumour cells were present.
Speed. About 1.4 s per hot-spot, against 10–15 minutes for exhaustive manual counting.
deep-SKIE. Tile accuracy on double stains was 98.4 / 90.9 / 91.0% (train / validation / test), against 95.9 / 86.9 / 84.8% on Ki-67-only stains.
Virtual double stains. SKIE graded all 6 hold-out WSIs correctly (index error 1.56 ± 1.15%), and deep-SKIE reached 87.1% accuracy on them.
Limitations noted by the authors
SKIE treats every nucleus inside synaptophysin-positive regions as tumour, so dense intratumoral inflammation contaminates the count.
H&E-to-double-stain registration uses manually chosen landmarks, although grade was robust to landmark variation in the cases tested.
G3 tumours were not studied, and double staining is not standard practice.
Reading notes
My own notes, not content taken from the paper.
This is a textbook case of the gold standard paradox. The "gold standard" pathologist chose worse hot-spots than the algorithm, so a naive agreement statistic counts the algorithm's best calls as errors. Only re-examining the discordant cases showed which side was right.
On the lecture slide
From a lecture slide titled "Top quantitative models in GI/liver path":
GI neuroendocrine tumor
SKIE / deep-SKIE
Ki-67 index
42 cases
Govind et al. Sci Rep. 2020
"42 cases" matches deep-SKIE's training and validation set (and the cycle GAN's training WSIs). SKIE itself was evaluated on all 50 cases.