About the Usage of Digital Pathology
Start with the intended use
Before choosing equipment or a viewer, describe what the service will do: teaching, research, consultation, or primary diagnosis; which specimens and stains it will cover; who will use it; and where interpretation will take place. Record what is outside scope and when to return to the glass slide.
This is an educational planning guide. The checklist is a practical synthesis, not a laboratory SOP or a statement that a particular installation is validated. Local clinical, information-governance, regulatory, and accreditation requirements need separate assessment.
Plan the complete workflow
Slide preparation and identification
Does the slide meet the laboratory's preparation criteria? Can case, specimen, slide, and image be matched without ambiguity?
Identity checks and rejection/rescan reasons
Scanning and image QC
Is all relevant tissue captured and in focus? Are there stitching errors, missing regions, or other artifacts? Are the expected images present?
Scan settings, QC outcome, and corrective action
Viewing and interpretation
Are the viewer, display, environment, network, and user training suitable for the intended work?
Configuration, training record, and known limitations
Storage and retrieval
Can authorised users retrieve the correct image and associated information promptly? Who owns capacity planning, backup/restore testing, retention, and support?
Storage and access plan, restore-test and maintenance records
Exceptions and downtime
What happens when an image is incomplete, a case is unsuitable, or a system is unavailable?
Escalation, glass-slide fallback, and incident procedures
The NPIC Guide to Digital Pathology provides implementation guidance on image quality, case matching, infrastructure, and ongoing costs. Backup tests, retention, and downtime arrangements above are planning prompts; this page does not prescribe a universal interval or retention period.
Validate diagnostic use locally
The active CAP/ASCP/API whole-slide imaging guideline is the 2021 update, published in print in 2022. Its recommendations and good-practice statements include:
At least 60 cases, representative of one intended application and routine diagnostic complexity; an additional 20 cases for relevant additional applications not represented in the initial set.
Comparison of the same observer's digital and glass-slide diagnoses, with at least a two-week washout between modalities.
Investigation and remediation when intraobserver concordance is below 95%. This is an investigation trigger, not a universal regulatory pass/fail mark or a claim of 95% diagnostic accuracy.
Assessment of the complete system in its intended setting, trained users, representation of tissue in the digital image, and documentation of the study and approval.
The guideline's recommendations are not themselves mandatory, as CAP's FAQ explains. Check applicable regulatory and accreditation requirements separately. The CAP teaching presentation provides context; this short note does not replace the full guideline.
Keep validation, routine QC, and change control distinct
Initial validation addresses the intended use of the workflow. Routine QC checks whether today's images and service remain fit for that use. Change control determines what evidence is needed after a change. Maintain procedures for changes that could affect clinical results; do not treat every software update or replacement as identical. CAP's change-control examples distinguish, for example, a new scanner make/model from replacement by the same make/model. Record the impact assessment and any verification or revalidation needed.
Check the specific product and configuration
Regulatory status is tied to a device, its labeling, intended use, and compatible configuration. FDA records illustrate why authorization of one WSI device cannot be extended to every scanner, viewer, or research algorithm: the Philips PIPS decision specifies a system and use, while the Novo decision covers viewer software with specified compatible equipment. These are examples, not an exhaustive current product list. Check the current record and labeling for the actual product and jurisdiction before clinical use.
A research tool's availability, open code, or publication is a separate question from its suitability and authorization for a diagnostic workflow. See the software chooser for research-oriented starting points.
Test interoperability and sharing explicitly
DICOM whole-slide imaging supports tiled, multiresolution images and associated metadata. Compare conformance statements and test the required scanner–archive–viewer exchange: the DICOM conformance guidance makes clear that conformance alone does not guarantee interoperability.
Before sharing research or teaching images, inspect metadata, filenames, associated label/macro images, and identifiers embedded in pixels. Removing selected metadata is not sufficient evidence of de-identification. The DICOM confidentiality profile distinguishes metadata handling from pixel-data cleaning. Use the institution's approved sharing process; a technical cleaning step alone does not establish legal compliance.
Foundational reading
Whole Slide Imaging Versus Microscopy for Primary Diagnosis in Surgical Pathology: A Multicenter Blinded Randomized Noninferiority Study of 1992 Cases (Pivotal Study). American Journal of Surgical Pathology (2018). DOI: 10.1097/PAS.0000000000000948. Read its study population, exclusions, system, comparator, and endpoints before applying its findings to another setting.
The Gold Standard Paradox in Digital Image Analysis: Manual Versus Automated Scoring as Ground Truth. Archives of Pathology & Laboratory Medicine (2017). DOI: 10.5858/arpa.2016-0386-RA. Background for evaluating the reference used to judge an automated measurement.
Source check: 1 October 2026, for the guidance and official documents linked on this page. This is a source-based content update, not clinical review or validation of any laboratory, product, or analysis pipeline.